LITE Level 1 Instrument Settings Overview

How do I use the LITE header parameters?
Check the Operation Mode Parameters to identify:
  • datatake mode (day or night) - datatakemode
  • type of operational mode (normal, multiscatter, gain change, or in transition) - specialopsmode
The Modes of Operation Document summarizes, by orbit, the instrument settings that were used to configure LITE for each mode of operation.
Check the Validity Parameter to determine if the profiles are valid, questionable, or invalid.

Profiles may be questionable due to:

  • bad ISDB/AOE data,
  • boresight not aligned,
  • 355 or 532 nm high voltage greater than -1000,
  • 1064 nm high voltage greater than -399, or
  • underflow occurred in baseline subtraction region.
Profiles may be invalid due to:
  • motor/wheel/filter motion,
  • aperture wheel in closed position,
  • high voltage disabled, or
  • the number of saturated counts in a profile is greater than 460,
  • underflow occurred in the baseline subtraction region (see: outofrangsubreg355, outofrangsubreg532, outofrangsubreg064)
Check the Instrument Status Parameters to determine the status of the instrument.
Check the following Instrument Setting Parameters for each wavelength:
  • Surface Mode Parameters
  • Attenuation Parameters
  • Background Value Parameters
  • High Voltage Parameters (Measured voltages and Commanded voltages)
  • Energy Monitor Parameters
  • Filter Status Parameters
Check the Out of Range Conditions to determine if underflow or overflow conditions have occurred. The user will need to decide if out-of-range conditions cause the profiles to be invalid.